SK03370
Gamma compensated neutron ionization chamber (LND 50348 with MI Cables)
General Specifications
| MAXIMUM LENGTH (INCH/MM) | 186/4724.4 | 
| MAXIMUM DIAMETER (INCH/MM) | 2.0/50.8 | 
| EFFECTIVE LENGTH (INCH/MM) | 9.0/228.6 | 
| CATHODE MATERIAL (INTERNAL/EXTERNAL) | Aluminum | 
| FILL GAS | Nitrogen | 
| FILL PRESSURE (TORR) | 760 | 
| EFFECTIVE VOLUME (CM³) | 225.0 | 
| OUTER CASING MATERIAL | Aluminum | 
| ANODE CONNECTOR | MHV at end of MI Cable | 
| CATHODE CONNECTOR | MHV at end of MI Cable | 
| GUARD CONNECTOR | MHV at end of MI Cable | 
Operating Characteristics
| RECOMMENDED OPERATING VOLTAGE (VOLTS) | 500 | 
| OPERATING TEMPERATURE RANGE °C | -50 to +100 | 
| NEUTRON SENSITIVITY (AMPS/NV) | 4.0E-14 | 
| NEUTRON FLUX RANGE (NV) | 1E4 to 1E11 | 
| MAXIMUM OPERATING VOLTAGE (VOLTS) | 800 | 
| GAMMA SENSITIVITY (A/R/HR - Cs-137) | 1.0E-11 | 
| RESISTANCE - SIGNAL ELECTRODE TO CATHODE (OHMS) | >1.0E13 | 
| RESISTANCE - GUARD ELECTRODE TO CATHODE (OHMS) | >1.0E10 | 
Neutron Sensitive Material
| COATING MATERIAL | B10 | 
| ENRICHMENT (%) | 96 | 
| COATED THICKNESS (MG/CM²) | 1 | 
| COATED AREA (CM²) | 240 | 